Electromagnetic Emissions
DualWave Arthroscopy Pump
User’s Guide
LM1-0630-EN_E
Page 41 of 43
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Surge
IEC 61000-4-5
± 1 kV differential
mode
± 2 kV common
mode
± 1 kV differential
mode
± 2 kV common
mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
40% UT
(60 % dip in UT)
for 5 cycles
70% UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
95% Reduction
(10ms)
60% Reduction
(100ms)
30% Reduction
(500ms)
95% Reduction
(5s)
Mains power quality should be that of a typical
commercial or hospital environment. If the AR-
6480 DualWave Arthroscopy Pump requires
continued operation during power mains
interruptions, it is recommended that the AR-6480
DualWave Arthroscopy Pump be powered from
an uninterruptible power supply.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m
3 A/m @ 50 & 60
Hz
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
Note: UT is the AC mains voltage prior to application of the test level.