Setup
DualWave Arthroscopy Pump
User’s Guide
LM1-0630-EN_E
Page 18 of 43
Setup
4.0
4.1
How to Set Up the Console
Users are encouraged to contact their Arthrex representative if they require a
more comprehensive surgical technique.
NOTE: To minimize the effects of hydrostatic pressure differences on the actual
joint pressure, both the pump and joint must be in the same horizontal
plane.
4.2
AC Power Safety Considerations
The AR-6480 is powered by a medically rated universal AC input switching
power supply. This power supply allows users to connect the console to any
local AC mains outlet. Please use the appropriate plug and a reliable ground
conductor.
Arthrex supplies separate power cords for the U.S. and Germany (CEE 7/7) with
the AR-6480. Contact your Arthrex representative if you need a power cord that
must meet the electrical standards of another country.
NOTE: Extension cords must meet local electrical standards.
The console is designed to meet power-saving guidelines. The console has an AC
mains switch on the front panel [3]. When the AC mains switch is OFF, no
electrical power is drawn by the console.
When the AC mains switch is ON, the console executes a series of self-diagnostic
tests. Upon successful completion of these tests, the console displays on the
TPVD [5] the name and model number, Arthrex DUALWAVE. If the tests
discover a problem, an error message shows on the TPVD. Refer to Table 3 for a
complete list of TPVD Messages.
In the event of an AC power interruption, the console can run continuously
without fault for up to 10 milliseconds. If an AC power failure lasts longer than
10 milliseconds, the system will reset to default settings when AC power is
restored.
W A R N I N G !
Do not have device in direct contact with patient if high-frequency devices are in
use, or defibrillation of the patient is required.