Product Description
DualWave Arthroscopy Pump
User’s Guide
LM1-0630-EN_E
Page 8 of 43
AR-6480 Display Messages
2.2.3
The console’s Touch Panel Visual Display (TPVD) [5] provides information about
the status of the AR-6480 and the pressure and flow settings in real time. Table 3
describes each message and button when pump is in ready state.
Table 3
Touch Panel Visual Display Messages AND Iconography
Message
Cause
Explanation
Arthrex DualWave
Message appears when the AC mains power
switch is activated.
Power on message display.
** Tubing Out **
Message appears when tubing is not plugged
into the Tubing Sensor Coupler [2].
Check tubing installation.
** Door Open **
Message appears when the roller housing door
[12] or [9] is open.
Roller housing door is not
closed.
** Over Pressure **
Message appears when the sensed pressure
exceeds over-pressure software limit of 300
mmHg.
Software overpressure
condition.
Critical Failure
Message appears on the first line of the TPVD if
one of three conditions is met:
Failure Condition 1: ** Power Failure **
Appears if the power supply self-test fails when
the pump is turned on.
Failure Condition 2: ** OVP Detect Fail
*
*
Appears if the hardware overpressure diagnostic
test fails when the pump is turned on.
Failure Condition 3: ** Sensor Failure **
Appears if the pump detects a problem with the
pressure sensors.
Critical failure, cannot
continue operation.
** Power Failure **
Message appears if the power supply self-test
fails when the pump is turned on.
Power supply test fails.
** OVP Detect Fail **
Message appears if the hardware overpressure
diagnostic test fails when the pump is turned on.
Hardware overpressure
diagnostic fails.
** Sensor Failure **
Message appears if the pump detects a problem
with the pressure sensors.
Sensor failure.
** Pressure Fault **
Message appears when the pump is unable to
reach a desired set pressure within a specific
amount of time. This typically indicates an
improperly installed tubing set or a split in the
tube from continuous use.
Insufficient measured
pressure.
Remote Control Icon
Icon appears when the remote is attached.
Remote connected.
Foot Pedal Icon
Icon appears when the foot pedal is attached.
Foot pedal connected,
+
Button
The TPVD displays the pressure reading until
PRESSURE (+) button is pressed. Once pressed,
the displayed pressure reading will change to
the pressure setting. Each subsequent press of
the pressure buttons will increase the pressure
setting in increments of 5.
Pressure set increase,
Содержание AR-6480