Setup
DualWave Arthroscopy Pump
User’s Guide
LM1-0630-EN_E
Page 20 of 43
2.
Connect the receiver end of the power cord for the AR-6480 into the
AC socket [14] and the plug end to the facility AC mains supply.
3.
Connect the shaver power cord (If shaver detection is desired,
otherwise skip to step 5).
NOTE: For more on shaver detection, see 4.6 Shaver Detection.
4.
Power on the shaver system.
5.
Turn on the AR-6480 [3].
6.
Verify the status of the AR-6480 displayed in the TPVD [5].
7.
Connect the tubing in accordance with Section 4.7 or 4.8
8.
Close the roller housing doors.
9.
Attach the Remote Control Unit or Foot Pedal Unit, if applicable.
10.
Refer to Section 5.0, Operation, for specific information on how to
operate the AR-6480, including pressure and flow settings.
11.
Press the Run/Stop button [Table 3] to activate the pump motor.
4.6
Shaver Detection
The AR-6480 detects when a shaver handpiece has been activated. The shaver
detection feature operates differently, depending on if an AR-6430 Outflow Tube
Set is installed on the DualWave.
The Shaver detection feature depends on the differential of current draw from
the shaver system when it is powered on in a static state and when the shaver
handpiece is activated.
For correct operation of Shaver Detect:
1.
Insure Shaver Detect power cord is properly connected between the
Shaver console and the DualWave.
2.
Attach shaver handpiece and foot pedal (if used).
3.
Turn on Shaver Console.
4.
After Shaver Console sequence, turn on DualWave.
NOTE: Shaver detect feature will function as described in 4.6.1
– 4.6.2.
Shaver Detect, Inflow only
4.6.1
During inflow only (No outflow tube set) when the shaver is activated the
DualWave activates BOOST mode for increased pressure to compensate for the
loss of distention from the vacuum used with shaving device.
The pressure will automatically change to the preset BOOST values, and can be
noted on the TPVD. The BOOST button will change to a dark blue color when the
shaver is activated.
If the LAVAGE mode is ON prior to the shaver activation, the LAVAGE mode
will be overridden except when LAVAGE mode causes a larger pressure
increase.
Содержание AR-6480