Setup
DualWave Arthroscopy Pump
User’s Guide
LM1-0630-EN_E
Page 22 of 43
•
The Shave Console must be completely powered up before powering
on the pump.
•
A new baseline is captured when pressing the RUN button which
characterizes a NOT ACTIVATED condition. The screen will display
a message reading, “ACQUIRING SHAVER BASELINE”. During the
baseline capture, do not activate a shaver.
Cautions to User:
•
As the THRESHOLD value decreases, so does the DualWave’s ability
to prevent false detections of shaver activation. This condition makes
the DualWave behave like the shaver handpiece is activated when it
is not. This is characterized by sporadic movement of the pinch roller
between the Shaver and Cannula tube ends. Conversely, as the
THRESHOLD value increases, so does the likelihood of the DualWave
not recognizing shaver activation. This can result in failure of the
DualWave to provide for aspiration through the shaver blade and
result in retention of heat within the blade.
•
As the DELAY values decreases the DualWave will have less time to
analyze the information received through the shaver detect circuit.
This action can result in a false detection. Increasing the DELAY value
should alleviate this condition. As the DELAY value increases, the
accuracy of the determination of the shaver handpiece state will
improve, but at the highest levels, the DELAY time may result in a
perceived lag before the DualWave responds to the shaver activation.
•
The shaver detect function of the DualWave has been verified to
function properly with the Arthrex Adapteur II shaver console. It is
the responsibility of the user facility to evaluate 3rd party shaver
consoles to determine the correct operating parameters and any
potential hazards of utilizing the DualWave with an alternate shaver
console.
4.7
How to Set Up the Pump Tubing
NOTE: These instructions describe the procedure to set up the Main Pump
Tubing or the ReDeuce Pump Tubing.
1.
Remove the orange cap from the Pump Tubing and insert the
Pressure Line Connector [e] of the Pump Tubing into the Tubing
Sensor Coupler [2].
This step must be completed first to ensure
accurate pressure measurements
.
2.
Open the inflow door completely. Allow the door to rest against the
stop. The roller mechanism is now exposed.
3.
Place the green-collared section of the Pump Tubing [c] into the
Inflow Tubing Track [1] indicated by the green dot.