Appendix
75
Turn the knob exactly to ring position number 1 and press the
SET1
soft button to confirm and store the alignment position. Repeat the
procedure until you finish with position 16.
Press the
UNDO
soft button if you missed the correct alignment
position and repeat the last step.
Press the
READY
soft button at the end of the procedure to return
to the
SERVICE MODE
screen.
The new calibration becomes effective after rebooting the WCU-4.
Knob Setup screens