Zoom Calibration Mode
Press the
ZOOM
soft button to enter
the
Stick Setup
screen.
If the zoom control acts in the wrong
direction, move the zoom stick up-
wards and press ADJ-DIR.
If the zoom zero position is not cor-
rect (issue: the zoom motor moves
without pressing the stick), leave the
zoom stick in its center position (do
not touch it) and press ADJ-0.
Press
BACK
to return to the
SERVICE MODE
main screen. The
new calibration becomes effective af-
ter rebooting the WCU-4.
Knob Calibration Mode
Press the
KNOB
soft button to enter the
Knob Setup
screen.
Turn the knob clockwise to its end position and press
K-CW
. Then
turn the knob counter-clockwise to its end position and press
K-
CCW
.
Attach the WCU-4 Knob Calibration Ring K2.0000848.
Press
TEST
to check if the engravings on the Knob Calibration Ring
align with the corresponding values on the WCU-4 screen. If they
do not align, press the
K-LIN
soft button to start the knob linearizing
procedure.
NOTICE
Do not start the linearizing procedure without the ARRI
calibration ring applied. Once started, you must go through the
whole alignment procedure.
By pressing
K-LIN
all previously stored alignment points will be
cleared and cannot be restored!
Please note that previously hand-marked focus rings might have
an offset after re-calibrating the focus knob!