25
Always Follow these 17 Steps for Cleaning
and Disinfection
Removing visual residue (step 1-3 of 17)
1
Remove the head rest
(A)
, the thigh support cushion
(B)
, the
patient security grip
(C)
and the safety belt/belts, and place
them at the bottom of the bathtub.
(See Fig. 1)
2
Position the
Miranti
lift bath trolley over the bathtub and
lower it down.
(See Fig. 2)
3
Clean:
• the upper part of the
Miranti
lift bath trolley (start from top
and move downwards)
• all removed parts
from visible residue with a cloth soaked in water.
Alternatively shower with water using a brush and/or wipe
with a cloth.
Continue with the steps on the next page.