4 - 4
FULL FEATURE TEST
#
Item
Function
Activation/
Validation
Actions
Criteria for approval
1
Charging
Manual or button Push R/L button or direct lift
manually on a charging station
Activation of the charging station
For ECS option, system shall charge when
battery level is under 80%
2
Horizontal
displacement
Buttons
Push right and left function on
the hand control
No abnormal noise. Lift is moving along the rail
in the expected direction.
3
Vertical
displacement
Buttons
Push up and down function on
the hand control
No abnormal noise. Spreader bar is moving up
and down.
4
Inclined PDPS
Buttons
Push inclined or sitting position
No abnormal noise. Powered DPS is moving
according to the wanted action.
5
High / Low limit
switch
Buttons
Push up and down function on
the hand control
High limit switch must activate at top of stroke.
No abnormal noise. Low limit switch must
activate at end of stroke.
No tension limit must stop lowering action if
there is no tension on strap (e.g. without ac-
cessories)
6
Hand control
Buttons
Press each button
All functions are working: no display issues.
7
Up / Down
Buttons
Press alternate buttons on lift
Strap is moving up and down
INTRODUCTION
This document covers the maintenance and repair of the
Maxi Sky
2 and
Maxi Sky
2 Plus.
Maxi Sky
2 has a maximum capacity of:
- 272 kg (600 lb).
The
Maxi Sky
2 Plus has a maximum capacity of:
- 454 kg (1000 lb) when set in dual mode.
- 272 kg (600) when set in single mode.
FEATURES
THE SAFE WORKING LOAD (SWL) WILL DEPEND ON THE LIFT’S
CONFIGURATION MAXIMUM SWL REFERS TO THAT OF THE LOWEST RATED
ATTACHMENTS, AS WELL AS RAIL INSTALLATION
NOTE
The emergency features of the
Maxi Sky
2 ceiling lift include:
- Stop.
- Overload current limit protection in the PCB.
- Upper limit switch.
- Lower limit switch/no tension limit.
- Control panel on the unit / LED display
- Manual emergency lowering mechanism.
- Audible variable low-battery warning.
- Malfunction & maintenance code.
These tests will be performed after replacement of a part or a repair is performed to ensure the quality of the repair. The
test will be called out throughout section 6.
To perform a Full Featur Test on the MS2 Plus, refer to
Maxi Sky
2 Plus document “Initial Start Up” 001-31229-EN section
14 “Testing the
Maxi Sky
2 Plus System”.
001-15697 rev
. 5
Содержание Maxi Sky 2
Страница 1: ...Maxi Sky 2 Maintenance and Repair Manual with people in mind 001 15697 Rev 5 09 2015 ...
Страница 2: ...NOTE This page is left blank intentionally ...
Страница 6: ...NOTE This page is left blank intentionally ...
Страница 8: ...NOTE This page is left blank intentionally ...
Страница 10: ...2 2 NOTE This page is left blank intentionally 001 15697 rev 4 ...
Страница 12: ...3 2 NOTE This page is left blank intentionally 001 15697 rev 5 ...
Страница 23: ...4 11 NOTE This page is left blank intentionally 001 15697 rev 5 ...
Страница 112: ...7 11 MAXI SKY 2 PLUS C STAT WIRING DIAGRAM 001 15697 rev 5 ...
Страница 113: ...7 12 MAXI SKY 2 PLUS ECS WIRING DIAGRAM 001 15697 rev 5 ...
Страница 132: ...8 19 TROUBLESHOOTING 001 15697 rev 5 ...
Страница 133: ...8 20 001 15697 rev 5 ...
Страница 134: ...8 21 001 15697 rev 5 ...
Страница 135: ...8 22 001 15697 rev 5 ...
Страница 137: ...8 24 001 15697 rev 5 ...
Страница 138: ...8 25 001 15697 rev 5 The Troubleshooting section continues on the next page ...