-Using a 8 mm wrench unscrew the M5 bolt that holds the ring
terminal of the serial communication cable.
-Cut the tie wrap that holds the serial communication cable and
unplug it.
Unplug the swivel connection cable. Depending if the
Maxi Sky
2
Plus is ECS or C-Stat, the cable can be unplugged from two differ-
ent locations.
C-STAT Configuration
-For C-STAT configuration unplug the swivel connection cable from
the 12 pin connector using the pin extractor. Refer to section 6-31.
ECS Configuration
-For ECS configuration unplug the cable from the connector J2 on
main board.
-Pull out both the serial communication cable and swivel connection
cable from the ceiling lift enclosure.
001-15697 rev
. 5
Содержание Maxi Sky 2
Страница 1: ...Maxi Sky 2 Maintenance and Repair Manual with people in mind 001 15697 Rev 5 09 2015 ...
Страница 2: ...NOTE This page is left blank intentionally ...
Страница 6: ...NOTE This page is left blank intentionally ...
Страница 8: ...NOTE This page is left blank intentionally ...
Страница 10: ...2 2 NOTE This page is left blank intentionally 001 15697 rev 4 ...
Страница 12: ...3 2 NOTE This page is left blank intentionally 001 15697 rev 5 ...
Страница 23: ...4 11 NOTE This page is left blank intentionally 001 15697 rev 5 ...
Страница 112: ...7 11 MAXI SKY 2 PLUS C STAT WIRING DIAGRAM 001 15697 rev 5 ...
Страница 113: ...7 12 MAXI SKY 2 PLUS ECS WIRING DIAGRAM 001 15697 rev 5 ...
Страница 132: ...8 19 TROUBLESHOOTING 001 15697 rev 5 ...
Страница 133: ...8 20 001 15697 rev 5 ...
Страница 134: ...8 21 001 15697 rev 5 ...
Страница 135: ...8 22 001 15697 rev 5 ...
Страница 137: ...8 24 001 15697 rev 5 ...
Страница 138: ...8 25 001 15697 rev 5 The Troubleshooting section continues on the next page ...