8 - 26
001-15697 rev
. 5
For MS2 Plus a good reference for troubleshooting can be found in MS2 Plus Initial Start Up 001-31229-EN
Troubleshooting section.
For Malfunction 24:
- Ensure serial communications cable is properly connected to J8 and J16 of both ceiling lifts.
- If problem persists, replace serial communication cable.
For Malfunction 26:
- Ensure that SWL red jumper (272 kg) of both ceiling lifts are properly connected to connector J12.