7 - 5
TECHNICAL DATA
GENERAL
MAXI SKY
2
ELECTRICAL
OPERATION, TRANSPORT AND STORAGE CONDITIONS
Weight, complete (Four-function model)
12.3 kg (27 lb)
Weight, complete (Two-function model)
11.6 kg (26 lb)
Lifting capacity
120 kg (265 lb), 200 kg (440 lb), 272 kg (600 lb)
Strap length
2500 mm (98.4 in)
Lifting speed
5.1 cm/s (2.0 in/s) without load
4.2 cm/s (1.7 in/s) at 120 kg (265 lb)
3.8 cm/s (1.5 in/s) at 200 kg (440 lb)
3.3 cm/s (1.3 in/s) at 272 kg (600 lb)
Maximum stroke (from ceiling)
2300 mm (90.6 in)
Horizontal displacement speeds
10, 15 and 20 cm/s (4, 6 and 8 in/s)
Operating force of control
< 5 N
Normal Operation
Ambient temperature range: 10°C to 30°C (50°F to 86°F)
Relative humidity range: 10% to 70%
Atmospheric pressure range: 700 hPa to 1060 hPa
Storage
Ambient temperature range: -40°C to 70°C (14°F to 158°F)
Relative humidity range: 10% to 70%
Atmospheric pressure range: 500 hPa to 1060 hPa
WARNING: This equipment is not suitable in the presence of flammable anesthetic mixtures with air or
oxygen, or with nitrous oxide.
Duty cycle
Up/Down Motor: Max 10%, 1 minute continuously
Left/Right Motor: Max 20%, 5 minutes continuously
Sit/Reclined Motor: Max 10%, 2 minutes continuously
Horizontal axis motor
24 VDC, 62 watts
Vertical axis motor
24 VDC, 110 watts
Rating
24 VDC, 15 A max.
Noise level for either raising or lowering
61 dBA max. (with or without load)
Medical equipment
Type BF protection against electrical shock in accordance
with IEC 60601-1
WARNING: Wireless communications equipment such as wireless home network devices, mobile phones,
cordless telephones and their base stations, walkie-talkies, etc. can affect the Maxi Sky 2 and should be
kept at least 2.3 m away from it. Cables from potentially strong sources of electromagnetic fields should
not be placed near the unit.
Battery type
12 V, 5 Ah, sealed rechargeable valve regulated lead acid battery
Battery capacity
Provides up to 120 transfers with a load of 100 kg (220 lb), up to
70 transfers with a load of 200 kg (440 lb)
Degree of protection - Hand control
IPX7
Degree of protection -
Maxi Sky
2
IP24
Lift - protection class - shock prevention
Internally powered equipment
Battery Charger input
100-240 VAC, 50-60 Hz, 57 - 70 VA.
Battery Charger output
28.1 VDC, 1 A max, 28.1 VA
Battery Charger safety protection
Class 2, double insulated
001-15697 rev
. 5
Содержание Maxi Sky 2
Страница 1: ...Maxi Sky 2 Maintenance and Repair Manual with people in mind 001 15697 Rev 5 09 2015 ...
Страница 2: ...NOTE This page is left blank intentionally ...
Страница 6: ...NOTE This page is left blank intentionally ...
Страница 8: ...NOTE This page is left blank intentionally ...
Страница 10: ...2 2 NOTE This page is left blank intentionally 001 15697 rev 4 ...
Страница 12: ...3 2 NOTE This page is left blank intentionally 001 15697 rev 5 ...
Страница 23: ...4 11 NOTE This page is left blank intentionally 001 15697 rev 5 ...
Страница 112: ...7 11 MAXI SKY 2 PLUS C STAT WIRING DIAGRAM 001 15697 rev 5 ...
Страница 113: ...7 12 MAXI SKY 2 PLUS ECS WIRING DIAGRAM 001 15697 rev 5 ...
Страница 132: ...8 19 TROUBLESHOOTING 001 15697 rev 5 ...
Страница 133: ...8 20 001 15697 rev 5 ...
Страница 134: ...8 21 001 15697 rev 5 ...
Страница 135: ...8 22 001 15697 rev 5 ...
Страница 137: ...8 24 001 15697 rev 5 ...
Страница 138: ...8 25 001 15697 rev 5 The Troubleshooting section continues on the next page ...