8. Troubleshooting and Alarm Conditions
The following table provides a troubleshooting and alarm condition guide for the
Alpha Active
4
system in the event of malfunction. These alarms do not cause any
delay or interruption in therapy.
Indicator
Possible Cause
Remedy
LOW PRESSURE
The tubeset is not
connected properly.
CPR not fully closed.
There is a leak in the
system.
Check the tubeset
connector and ensure
it is securely fitted to
the pump.
Close CPR unit.
Call service engineer.
Low priority according
to IEC60601-1-8
POWER FAILURE
Power has been
removed from the
pump.
Re-apply power or
switch the pump off.
Low priority according
to IEC60601-1-8
SERVICE
Pump has detected an
internal fault, such as
the gearbox failure.
Switch the pump off
and call service
engineer.
Low priority according
to IEC60601-1-8