SH-7000-Pre-E-04/16
6
2
Set
up
M
ode:
To carry out calibrations, adjustments or data transfers, the pump has to be started in
setup mode. In the following chapter this process is described in detail.
2.1
Starting the pump in Setup Mode
1.
press and hold right down arrow “V” key on the “VOLUME ML” window
2.
press “ON/OFF”
3.
enter 1111 in the “VOLUME ML” window with the arrow “V /
Λ
”
keys
4.
confirm with “MENU/SELECT”
5.
“RATE ML/H” and “VOLUME ML” window show 0
You have now successfully entered the setup mode of the Volumed®
µVP7000. The
address number is displayed in the “RATE ML/H” window. In the “VOLUME ML” window
the current values on the corresponding address is shown. The values can be changed
with the arrow “V /
Λ
” keys.
Figure: Entering the password for Setup Mode (1111)
Each parameter and its functionality are described in the chapters 2.3 and 2.4.
2.2
Description of pump setup.
The database of the pump consists of approximately 55 addresses. They consist of the
following groups:
Address 0-19 = Basic settings, calibration
Address 20-55 = Optional settings
Basic settings are superior parameters, which define the actual over all calibration of the
device. With the optional settings the LCD contrast or the language etc. can be changed.
Manual changes on the option settings are only possible when address 50 = 0. The
optional setting for each configuration can only be changed with arcomed ag own
PremiumLink™ software. For the usage of PremiumLink™ separate trainings are highly
recommended. Please contact arcomed ag for further details.