SH-7000-Pre-E-04/16
43
7
Calibration
The pump must calibrated with the final user IV set. Only one brand of IV set is allowed.
All calibrations must be done with the pump on configuration 0.
To start up the pump in Setup Mode, please refer to chapter
2.
7.1
Pressure Sensor
The built in pressure sensor of the Volumed® µVP7000 Chroma has to be adjusted first
mechanically and then by means of software. The mechanical adjustment is necessary
after replacement or if the software adjustment is not sufficient.
Used tools:
- 2.5 mm Hex driver
- 0,89 mm Hex driver
- Infusion giving set filled with water
- Manometer with three ways tab. A syringe filled with water must be connected to the
manometer.
Procedure:
1.
Turn off pump and unplug open housing according to instructions for changing the
Main PCB (6.2.1).
2.
Dismantle the Pump block (6.2.4)