Maintenance
Maintenance and safety checks have to be done according to the checklist.
Worn or defective parts must be replaced. However, the pumps do not need any regular
replacement of specific items. arcomed ag provides service kits and it might be more efficient to
use all parts provided in the kits to avoid additional repairs. It is however not mandatory to use all
parts in the kits.
8.1
Scope and Schedule of Safety Checks
Schedule: 24 months or 10,000 hours of operation,
Technical staff that have been trained and authorised in writing by
arcomed ag
or Arcomedical
Infusion Ltd must check the unit according to the following list:
Visual check
Check for
Result
Housing, front foil
Drop detector
Door, hinges,
Stop-flow-mechanism
Cover of the peristaltic mechanism
Readability of the superscripts
Indicators (LED, etc.)
Mains connection, fuse
Damages
Completeness
Cracks
Usage
Cleanness
Function
!
ok