SH-7000-Pre-E-04/16
39
6.2
SW Update
!!WARNING!!:
Pump must be on
mains!
1.
Turn pump off and connected to mains power
2.
Connect USB interface to PC, start PremiumLink™, open set or import .BIN file
3.
Select
“Update SW”
,
“Pump Text”
and
“Auto”
in the download to pump section
4.
Position IR windows of interface and pump over each other. Ideal is the
UniQueDoc
™
Rack Module with adapter for IR interface
!!CAUTION!!:
Do not use upper
IR-window for data transfer! Min. distance
between
IR-windows is
10mm
! Pump and IR interface have to stand stable.
Do not interrupt
data transfer! Always press “Download”
1x
in PL. It is recommended to use a mouse.
5.
Press
“Download” once
in the download to pump section
.
Pump will go into
updating automatically. First the LCD will blink, after the buzzer will give clicking
sounds. Wait until the pump goes back into normal off mode.
6.
Turn pump on and off again.
7.
Un-select
“Update SW”
and
“Pump Text”.
Select
“Auto”
,
“Drug Modes”
,
“Configuration”
and
“Set Time”
in the download to pump section
8.
Press
“Download” once
in the download to pump section
.
Pump will go into
updating automatically. First the LCD will blink, after the buzzer will give clicking
sounds. Wait until the pump goes back into normal off mode.
9.
In order to ensure the proper operation of the pump, perform your operation and
safety tests.
10. Update complete