Functional test
Check for
Result
1 Stop-flow-mechanism
Open door, insert primed infusion set. Close the
door and turn on the pump. An automatic test of
the Stop flow mechanism will start. After 10
seconds the pump must ask for entering the rate
and VTBI.
!
ok
2. Pressure
Open the door, fill IV set with water, prime the set,
place set in pump and connect to the manometer.
Switch on the pump and build up a pressure of 500
mbar with the syringe (read out on the
manometer). Go to the occlusion pressure menu
The displayed pressure has to be within 350 and
650 mbar.
p =
3. Door pressure check
Insert infusion set filled with water, into pump, turn
pump on wait for the stop flow test to be
completed. Connect infusion set to a mano-meter.
Start pump in Setup Mode and go to Address 15.
Built up pressure of 800 mbar externally. Press and
hold the “START” key. Watch the manometer, the
pressure should not fall below 700 mbar at the
beginning. Keep pressing the “START” key and
watch how the pressure rises.
PVC set: maximum pressure (pmax) has to be
within 1.5 bar and 2.5 bar.
Silicon Set: maximum pressure (pmax) has to be
within 1.2 bar and 1.5 bar
pmin =
pmax =
4. Volume test
Insert filled set and turn on the pump. Set rate to
400 ml/h, volume to 10 ml. Start the pump and
check volume with a scale or similar device. The
cumulated volume must be between 9,7 and 10,3
ml.
V =
5. Test run
Rate: 100 ml/h, 30 min
PVC set: precision: max. +/- 5 %
Silicon set: precision: max. +/- 3 %
!
ok
6. Occlusion test
Insert filled set into the pump and start it with a rate
of 100 ml/h. Kinking the tube on the patient side
simulates an occlusion. After a short while, the
occlusion alarm should come on.
!
ok