![Arbiter Systems Portable Power Sentinel 933A Скачать руководство пользователя страница 64](http://html1.mh-extra.com/html/arbiter-systems/portable-power-sentinel-933a/portable-power-sentinel-933a_operation-manual_2962845064.webp)
3.9.2
Configuring
Before using the current module, make sure to press the SETUP key, then
6
−
Current Setup to configure the range. You should see the screen/menu
similar to that shown in Figure 3.11.
Notice from the table in screen shot that there are two ranges for
current and two ranges for voltage. The two ranges for current are 50 mA
and 1.2 Amp. The two ranges for voltage are 100 mV and 2.4 Volts.
Soft key Selections
1. Make sure to press the MODIFY RANGE soft key to select the
desired current or voltage input range for each channel.
2. Press the MODIFY INPUT soft key and select either current or
voltage corresponding to the input type.
3. Before leaving this menu, press the SAVE SETTINGS soft key. To
leave without saving press the EXIT soft key.
3.10
Trigger Setup
The last selection in the setup selection screen is called “Trigger Setup.”
Triggers allow the Model 933A to automatically record to internal flash
specific data when either the measured input signal, or calculated DSP
value, meets the defined conditions.
The example Trigger Setup shown below illustrates the configured fields
to trigger an event when the A-phase voltage exceeds 130 Vrms.
For
complete details on setting up triggers, go to Chapter 7, Triggers and
Event Recording.
Figure 3.12: Trigger
Review Menu
46
Содержание Portable Power Sentinel 933A
Страница 16: ...B 3 Uploader Progress Window 165 B 4 Successfully Uploaded File Information 166 xvi ...
Страница 18: ...xviii ...
Страница 35: ...Figure 2 2 Front Panel Description 17 ...
Страница 128: ...Figure 7 9 Power Quality Reference Selection Softkey Figure 7 10 Configuring the Power Quality Reference 110 ...
Страница 163: ...Chapter 10 Connection Diagrams 145 ...