Figure 6.6: Configure Measurement Parameters, DSP Mode
Loss Compensation
Loss Compensation allows you to compensate for the iron and copper losses
(real and imaginary) when connecting through a transformer. Compensate
for transformer loss only through PSCSV, not through 933A front panel.
Anticreep
Anticreep sets a threshold below which the 1133A will not accumulate
energy in the specific registers. This is similar to restricting an older style
(electromechanical) meter from spinning when the energy value was below
a certain threshold. Older electromechanical meters would spin due to the
fact that they were merely energized and energy was not consumed by the
user. Anticreep may only be configured from PSCSV software. First, set
to “ON” in DSP Mode, then set Anticreep.
6.2.7
Other Windows
Save As
When selecting to broadcast any data, it will open with a Save As window.
Since the data that you are broadcasting is ending up on your computer
(monitoring the 933A measurements), it needs a file name. The SaveAs
window provides a default file name. Choose this one, or you can type in
another, and click OK.
Limit File Size
So that the data file does not take unlimited memory on your computer,
the setting allows you to specify how much data you want to collect before
89
Содержание Portable Power Sentinel 933A
Страница 16: ...B 3 Uploader Progress Window 165 B 4 Successfully Uploaded File Information 166 xvi ...
Страница 18: ...xviii ...
Страница 35: ...Figure 2 2 Front Panel Description 17 ...
Страница 128: ...Figure 7 9 Power Quality Reference Selection Softkey Figure 7 10 Configuring the Power Quality Reference 110 ...
Страница 163: ...Chapter 10 Connection Diagrams 145 ...