Aquaprobe
®
AP-7000 Instruction Manual
10401-00894 Rev M
15.4.6.
Turbidity Zero Point Calibration
To calibrate the Turbidity zero point (zero the electrode), follow these steps:
1. Pour 400mL of clean water (bottled still mineral water is recommended) into a clean
calibration tube, remove the storage cap from the pH electrode if fitted, wash the
Probe in clean water, then drop the Probe in all the way.
The Sleeve End Cap must
be fitted
.
2. Switch the Aquameter
®
on and wait until the temperature and turbidity readings are
stable. Activate the probe cleaning feature in order to remove any air bubbles that
may be clinging to the electrodes.
3. Ensure the temperature of the solution is between 5ºC and 40ºC (41ºF - 104ºF).
4. Press the
MENU
key then select
Calibration.
The following screen will be displayed.
Calibration
RapidCal
DO 100%
Full Cal
5.
Select
Full Cal.
The screen will change to:
Calibration
pH/REDOX(ORP)
DO/EC
Aux Electrodes
6.
Select
Aux Electrodes
. The screen will change to:
SELECT ELECTRODE
1:TURB
l
4:EMPTY
2:EMPTY
l
5:EMPTY
3:EMPTY
l
6:EMPTY
Select the TURB electrode. The screen will change to:
CALIBRATE TURB
ZERO? [01/Jan/14]
1000? [01/Jan/14]
20? [01/Jan/14]
The dates shown to the right of each point are the dates of the last successful calibration.
7.
Select ZERO. The screen will change to:
PLEASE WAIT
Stabilising
000%