Aquaprobe
®
AP-7000 Instruction Manual
10401-00894 Rev M
4. Remove the new Optical DO Cap from its light-proof bag and quickly screw it onto
the end of the DO/EC electrode. Ensure that the cap is screwed fully onto the
electrode and that it is done up tight.
5. Carry out both Zero point and 100% point DO calibration as described earlier.
Please Note: It is essential when replacing the Optical DO Cap to calibrate the Zero
point BEFORE calibrating the 100% point.