Aquaprobe
®
AP-7000 Instruction Manual
10401-00894 Rev M
7.
Select
DO/EC
. The screen will change to:
Calibration
DO Zero?[01/Jan/17]
DO 100%?[01/Jan/17]
EC R-CAL?[01/Jan/17]
The dates shown to the right of the screen are the dates of the last successful calibration.
The value shown on the bottom line next to ‘EC’ is the value the EC electrode was last
calibrated to.
8.
Move the pointer down to the bottom line using the down arrow key.
Calibration
DO Zero?[01/Jan/17]
DO 100%?[01/Jan/17]
EC R-CAL?[01/Jan/17]
If the Calibration Standard value you are using is already displayed, press the
OK
key to
start calibrating. Remember, if you are using RapidCal solution, the EC value on this line
should be R-CAL.
If the value of the EC Calibration Standard you are using is not displayed, press the right
arrow key. The bottom line will change to:
Calibration
DO Zero?[01/Jan/17]
DO 100%?[01/Jan/17]
EC
R-CAL?[01/Jan/17]
You can now use the up and down arrow keys to select one of two pre-set EC Calibration
Standard values (R-CAL or SC-35) or to input any calibration standard value between
100µS/cm and 99,999µS/cm.
The value you input should be the calibration solution's
EC value at 25ºC.
9.
Once the correct Calibration Standard value is being displayed, press the
OK
key. The screen will change to:
PLEASE WAIT
Stabilising
000%
10.
If the calibration is successful, the counter will reach 100% and the following
screen will be displayed.
Cell constant:0.98
Calibrating
100%
Press [OK]