![Aquaread Aquaprobe AP-7000 Скачать руководство пользователя страница 101](http://html.mh-extra.com/html/aquaread/aquaprobe-ap-7000/aquaprobe-ap-7000_instruction-manual_2960645101.webp)
Aquaprobe
®
AP-7000 Instruction Manual
10401-00894 Rev M
The AP-7000's sleeve and wiper should also be removed in order to reduce the Probe's
thermal mass. To calibrate the ISE electrode follow these steps:
Point 1.
1. Remove the Probe Sleeve and the Wiper (see section
). Remove the storage
cap from the pH electrode, wash the Probe in distilled water, dry the probe
thoroughly then gently lower the Probe in to the warm
10ppm
solution.
2. Switch the Aquameter
®
on and stir the probe until the temperature and NO3
readings are completely stable. A minimum of five minutes is recommended.
3. Ensure the temperature of the solution is between 20ºC and 40ºC (68ºF - 104ºF).
4. Press the
MENU
key then select
Calibration.
The following screen will be displayed.
Calibration
RapidCal
DO 100%
Full Cal
5. Select
Full Cal.
The screen will change to:
Calibration
pH/REDOX(ORP)
DO/EC
Aux Electrodes
6. Select
Aux Electrodes
. The screen will change to:
SELECT ELECTRODE
1:TURB
l
4:EMPTY
2:NO3
l
5:EMPTY
3:EMPTY
l
6:EMPTY
The Nitrate (NO3) electrode should have been assigned to an AUX socket when it was
fitted. Move the pointer to NO3 then press the OK or right arrow key to select.
7.The screen will change to:
CALIBRATE NO3
Pt-1? [01/Jan/17]
Pt-2? [01/Jan/17]
Pt-3? [01/Jan/17]
Calibration point 1 (Pt-1) is the warm 10ppm point. Calibration point 2 (Pt-2) is the warm
100ppm point. Calibration point 3 (Pt-3) is the cool 10ppm point.
The dates shown to the right of each point are the dates of the last successful calibration.
© 2017 Aquaread
®
Ltd.
www.aquaread.com
Page 101 of 143