![Aquaread Aquaprobe AP-7000 Скачать руководство пользователя страница 43](http://html.mh-extra.com/html/aquaread/aquaprobe-ap-7000/aquaprobe-ap-7000_instruction-manual_2960645043.webp)
Aquaprobe
®
AP-7000 Instruction Manual
10401-00894 Rev M
13.7. Servicing the pH Electrode
1.
Remove the pH or combined pH/ORP electrode from the Probe body
(see
Electrode Removal and Replacement
2.
Rinse with methyl alcohol.
3.
Replace the electrode.
4.
Re-calibrate.
Never place the entire AP-7000 in methyl alcohol, as this will cause irreparable
damage to the DO/EC electrode. Damaged caused in this way is not covered by the
warranty.
If the methyl alcohol rinse does not restore the electrode, perform the following actions:
1.
Remove the electrode from the body again.
2.
Soak in 0.1M HCl for 5 minutes.
3.
Rinse in de-ionised water.
4.
Soak in 0.1M NaOH for 5 minutes.
5.
Rinse in de-ionised water.
6.
Soak in pH4.01 buffer for 10 minutes.
If the above procedure still does not restore performance, replace the electrode.
13.8. Calibrating ORP
ORP electrodes should be calibrated at least once a month to ensure optimum accuracy.
Full calibration involves calibrating at a single point, either +250mV (at 25
C) using a
+250mV ORP calibration standard such as
Reagecon RS250 Redox Standard,
or
+229mV (at 25
C) using a +229mV ORP calibration standard such as Zobell Solution.
For best results, calibrate as close to 25
C as possible. The probe will automatically
compensate for temperature variation in the calibration solution during calibration.
To calibrate the ORP electrode follow these steps:
1.
Fill a calibration bottle with fresh calibration solution, remove
the storage cap from the pH/ORP electrode, wash the Probe in distilled
water, then drop the Probe in all the way.
2.
Switch the Aquameter
®
on and wait until the temperature and ORP
measurements are completely stable.
3.
Ensure the temperature of the solution is between 5ºC and 40ºC (41ºF -
104ºF).
4.
Press the
MENU
key then select
Calibration.
The following screen will be
displayed.
Calibration
RapidCal
DO 100%
Full Cal
© 2017 Aquaread
®
Ltd.
www.aquaread.com
Page 43 of 143