Aquaprobe
®
AP-7000 Instruction Manual
10401-00894 Rev M
If there is no water flow across an ISE electrode, the ions in the immediate area of the
electrode will be depleted and the readings will start to fall.
During long term deployment of ion selective electrodes, the user should obtain grab
samples during the course of the deployment for analysis in the laboratory by chemical
means and use the results to apply post calibration to the recorded results.
During long term deployment, the AP-7000 can be suspended by the
Extension Cable, but
in order to provide a fail-safe for the plug and socket, always Ty-wrap the Extension Cable
to the bracket just above the plug.