Aquaprobe
®
AP-7000 Instruction Manual
10401-00894 Rev M
5. Connecting an AP-7000
The AP-7000 is designed to connect to the Aquameter
®
using an AP-7000 Extension
Cable. The AP-7000 Extension Cable features high-pressure metal connectors, which
incorporate several O-ring seals at the Probe end. Prior to first connection, the seals must
be lubricated using the silicone grease supplied.
Apply a generous smear of grease to the O-rings where indicated above. Be careful not to
get any grease inside the connector near the gold contacts. A small smear of grease
should also be applied to the thread on the Probe to allow easy tightening of the collar.
To connect the Extension Cable to the AP-7000, align the coloured dot on the AP-7000
with the
◄AQUAREAD
logo on the plug body, then press the plug into the socket and
tighten the retaining collar fully. Once the AP-7000 has been connected to the Extension
Cable, the Aquameter
®
can be connected.
Always ensure the Aquameter
®
is switched off prior to connecting or disconnecting an AP-
7000. Align the
◄AQUAREAD
logo on the plug body with the red on/off switch on the
Aquameter
®
, then press the plug into the socket and tighten the retaining collar.
Once the AP-7000 is connected to the Aquameter
®
, switch the Meter on by pressing the
red on/off switch. The Aquameter
®
should detect the Probe and start displaying readings.
5.1. AP-7000 Deployment
When you are ready to deploy the AP-7000, remove the mounting nut, fit the mounting
bracket then replace and tighten the mounting nut. During deployment, the AP-7000 can be
suspended by the Extension Cable, but in order to provide strain relief for the plug and
socket, always Ty-wrap the Extension Cable to the bracket just above the plug.