Aquaprobe
®
AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev R
The Meter will wait until the readings are stable, then it will send the calibration command
to the Probe, where the calibration takes place. During calibration, the Calibrating screen is
displayed and the progress counter counts up. If the calibration is successful, the counter
will reach 100% and the following screen will be displayed.
Output:348mV
Calibrating
100%
Press [OK]
The Calibration Report on the top line displays the voltage output from the ISE electrode in
millivolts (mV). This value is stored in the Probe's memory and can be recalled at any time.
See section
10.7 Calibration Data Storage and Retrieval
. Press
OK
then
ESC
repeatedly
to return to normal reading mode
.
Point 2
1. Remove the probe from the 10ppm solution and wash thoroughly in deionised water.
Dry the probe then drop it into to the warm
100ppm
solution.
2. Stir the probe until the temperature and NH4 readings are completely stable. A
minimum of five minutes is recommended.
3.
Ensure the temperature of the solution is within 1ºC of the previous 10ppm
calibration point.
If the solution is warmer or cooler than this, calibration will fail.
4. Referring to steps 4-7 above, select Pt-2 and press OK.
The Meter will wait until the readings are stable, then it will send the calibration command
to the Probe, where the calibration takes place. During calibration, the Calibrating screen is
displayed and the progress counter counts up as shown above. If the calibration is
successful, the counter will reach 100% and the calibration report screen will be displayed.
If the temperature of the 100ppm solution is more than 1ºC different from the Pt-1
calibration temperature, an OUT OF TEMP RANGE calibration error will be reported. If this
happens, adjust the temperature and try again.
Point 3
1. Remove the probe from the 100ppm solution and wash thoroughly in deionised
water. Dry the probe then drop it into to the
cool 10ppm
solution.
2. Stir the probe until the temperature and NH4 readings are completely stable. A
minimum of five minutes is recommended.
3.
Ensure the temperature of the solution is at least 10ºC cooler than the
previous 100ppm calibration point.
If the solution is too warm, calibration will fail.
4. Referring to steps 4-7 above, select Pt-3 and press OK.
The Meter will wait until the readings are stable, then it will send the calibration command
to the Probe, where the calibration takes place. During calibration, the Calibrating screen is
displayed and the progress counter counts up as shown above. If the calibration is
successful, the counter will reach 100% and the calibration report screen will be displayed.
If the temperature of the cool 10ppm solution is less than 10ºC cooler than the Pt-1 and Pt-
2 calibration temperatures, an OUT OF TEMP RANGE calibration error will be reported. If
this happens, adjust the temperature and try again.
© 2017 Aquaread
®
Ltd.
www.aquaread.com
Page 95 of 141