Aquaprobe
®
AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev R
To calculate a GS Factor, first take measurements using the fully calibrated electrode.
Next, compare the average of these values with the average values derived by laboratory
analysis of grab samples from the same location. To do this, divide the average grab
sample value by the average electrode value. This will give you a GS Factor.
For example, your calibrated electrode gives an average output of 100 at a given location.
The analysis of grab samples from that location reveal an actual value of 125. So, 125
divided by 100 gives a GS Factor of 1.25.
This value should now be input on the bottom line of the electrode's calibration screen.
Once the GS Factor value has been input, the OK key should be hit to send the Factor to
the Probe.
Now that this GS Factor has been applied to the electrode, all future measurements will be
multiplied by 1.25 prior to being displayed.
In this way, the electrode has been corrected for the local conditions and species of
chlorophyll.
16.7.9.
Errors During Calibration
If a problem occurs during calibration, an error message will be displayed. Refer to
Calibration Error Messages
in section 10 for error message handling.
16.7.10.
Lens and Sleeve Maintenance
On a daily basis, the lenses on the electrode should be wiped over with a soft damp cloth.
Similarly, the inside of the Probe Sleeve and Sleeve Cap should be kept clean and free
from any deposits that may cause stray fluorescence.
Never use an abrasive cleaner on the inside of the Probe sleeve or cap as they have
been treated with a non-reflective coating which can be easily damaged.
The inside of
the sleeve should be wiped over with a soft damp cloth and
non-abrasive
detergent.
Always re-calibrate the zero point after cleaning the sleeve or lenses.
© 2017 Aquaread
®
Ltd.
www.aquaread.com
Page 73 of 141