Aquaprobe
®
AP-700, AP-800 & AP-2000 Instruction Manual
10401-00870 Rev R
Point 1.
1. Remove the storage cap from the pH electrode, wash the Probe in distilled water,
dry the probe thoroughly then drop the Probe in to the warm
10ppm
solution.
2. Switch the Aquameter
®
on and stir the probe until the temperature and Cl readings
are completely stable. A minimum of five minutes is recommended.
3. Ensure the temperature of the solution is between 20ºC and 40ºC (68ºF - 104ºF).
4. Press the
MENU
key then select
Calibration.
The following screen will be displayed.
Calibration
RapidCal
DO 100%
Full Cal
5. Select
Full Cal.
The screen will change to:
Calibration
pH/REDOX(ORP)
DO/EC
Aux Electrodes
6. Select
Aux Electrodes
. The screen will change to:
SELECT ELECTRODE
1:TURB
l
4:N/A
2:Cl
l
5:N/A
3:N/A
l
6:N/A
The Chloride (Cl) electrode should have been assigned to AUX socket 2 when it was fitted.
Move the pointer down to 2:Cl then press the OK or right arrow key to select.
7.The screen will change to:
CALIBRATE Cl
Pt-1? [01/Jan/17]
Pt-2? [01/Jan/17]
Pt-3? [01/Jan/17]
Calibration point 1 (Pt-1) is the warm 10ppm point. Calibration point 2 (Pt-2) is the warm
100ppm point. Calibration point 3 (Pt-3) is the cool 10ppm point.
The dates shown to the right of each point are the dates of the last successful calibration.
8. Select Pt-1. The screen will change to:
PLEASE WAIT
Stabilising
000%