Using a clean cloth or tissue paper, polish the gold contacts ensuring they are completely
clean. Carefully insert the electrode into the AUX socket and tighten firmly until the O-ring is
completely compressed.
Optical electrodes, when fitted, sit adjacent to the DO electrode. If the DO electrode has a
red warning label fitted (see photo below), the label must be removed in order to prevent
reflections.
24.2. Socket Assignment and Calibration
After installation, it is essential to connect the Aquaprobe
®
to an Aquameter
®
and assign the
new electrode type to the relevant AUX Socket. On the Aquameter
®
, press the MENU key,
then select Setup & Install followed by Socket Assignment. When the Socket Assignment
option has been selected, the following screen will be displayed.
SOCKET ASSIGNMENTS
1:EMPTY
l
4:N/A
2:EMPTY
l
5:N/A
3:N/A
l
6:N/A
Using the up and down arrow keys, select the AUX socket you wish to assign then move
the cursor to the right by pressing the right arrow key. When the cursor has moved to the
right of the AUX socket number, use the up and down arrow keys to select the appropriate
electrode type. The tables below show the available electrode options and the selection
that should be made on this screen:
AP-2000/5000 type Optical Electrodes (AUX1 only)
Electrode Part No.
Function
Aquameter
®
Selection
2000/5000-TURB
Turbidity
TURB
2000/5000-CPHYLL
Chlorophyll
Cphl
2000/5000-BGA-PC
Phycocyanin (Blue-Green Algae PC)
BGA-PC
2000/5000-BGA-PE
Phycoerythrin (Blue-Green Algae PE)
BGA-PE
2000/5000-RHOD
Rhodamine WT Dye
Rhod
2000/5000-FSCEIN
Fluorescein Dye
Fcein
2000/5000-REFOIL
Refined Oil
R-OIL
2000/5000-CDOM
CDOM/FDOM
CDOM
AP-5000/7000 type ISE Electrodes (AUX1 only)
Electrode Part No.
Function
Aquameter
®
Selection
5000/7000-AMM
Ammonium/Ammonia
NH4
5000/7000-CHL
Chloride
Cl
5000/7000-FLU
Fluoride
F
5000/7000-NIT
Nitrate
NO3
5000/7000-CAL
Calcium
Ca2