7.2.2.1 I Range High, AC Low Current Coefficient
1. Make sure the load in connected (switch is closed) and adjust the load setting for a current
of 1 Aac +/- 10%.
2. Highlight the Measured I field value, then enter the current reading from the DAM.
Once the password has been entered press the soft key to the right of ‘Current Measure’.
7.2.2.2 I Range High, AC High Current Coefficient
The next calibration point is for high AC current measurements. Adjust the load to 24 Aac.
Note:
Depending on the power model and max. current rating, the actual high current value
may be different that shown in this example. Always follow the on-screen prompted current
setting values.
When done, press the ‘Continue’ key to the next calibration coefficient.
96
Calibration
PROFESSIONAL INNOVATIVE BRANDING SERVICE
When done, press the [Continue] key to the next calibration coefficient.
7.2.2.3 I Range High, DC Low Current Coefficient
1. Adjust the load setting for a current of 0.5 Adc +/- 10%.
2. Highlight the Measured I field value, then enter the current reading from the DAM.
When done, press the ‘Continue’ key to the next calibration coefficient.
1. Adjust the load setting for a current of 24 Aac +/- 10%.
2. Highlight the Measured I field value, then enter the current reading from the DAM.
7.2.2.4 I Range High, DC High Current Coefficient
1. Adjust the load setting for a current of 16 Adc +/- 10%.
2. Highlight the Measured I field value, then enter the current reading from the DAM.
97
Calibration
PROFESSIONAL INNOVATIVE BRANDING SERVICE