2.1.3
Waveform Viewer Setting
This item is for viewing the currently selected waveform.
In the ‘Settings’ screen, press the ‘Waveform Preview’ soft key to display the currently selected
waveform. Press the [ESC] key to turn off the viewer.
5
.
2.1.4
Limitation Setting
The limitation soft key displays the user limit setting screen. User limits are useful to protect a
unit under test from operator error by setting upper and or lower parameter value limits for key
output parameters like voltage and frequency.
‘Vac’ range is 0~300V.
‘F’ range is 15~1200Hz.
‘Vdc(+)’ range is 0~424.2V, ‘Vdc(-)’ range is -424.2V~0.
For example, to set the maximum output AC voltage.
1. In the ‘Limitation’ screen, move the cursor to highlight ‘Vac =’ field, press the desired [0]~[9]
keys and then [Enter] key to confirm.
2. Or, press the ‘Set to Maximum’ soft key directly.
32
Menu Operation
PROFESSIONAL INNOVATIVE BRANDING SERVICE
33
User limits are independent of the selected voltage range so an upper Vac limit setting
higher than 150V will have no effect while on the 150V AC voltage range as the voltage
range limit supersedes the user limit.
Note
5.2.2
Menu Setting
The following soft keys are available in the
Menu screen to access these settings:
5
.
2.2.1
Config Setting
The Config screen controls the following
aspects of the power source:
Remote Control
System Setting
System Information
External Control
Protection
1. Remote Control
1.1 RS232
The Remote Control screen provides access to RS-232, RS-485, GPIB and LAN interface.
Baud Rate, available rates are: 9600, 19200, 38400, 57600, 115200.
Parity Type: NONE, ODD, EVEN.
Stop Bits: 1, 2.
For example, to set Baud Rate to 19200.
Config (Configuration)
LCD Setting
Calibration
Other
Menu Operation
PROFESSIONAL INNOVATIVE BRANDING SERVICE
Содержание SP-300 Series