DIGITAL CALIBRATION
Solder Iron Digital Temperature Calibration Example
●
The external temperature sensor displays 250 degrees.
●
The set temperature and displayed actual temperature of the
soldering iron is 300 degrees.
●
300 — 250 = 50. An additional adjustment of 50 degrees is
required. Upon entering calibration mode, the display shows “010”,
indicating a calibration number of 10 is already present.
●
Therefore 10+50= 60.
●
We adjust from “010” to “060” by pressing the up button of the
Soldering Iron Temperature Adjustment button (“4” from the
control panel)
●
Save and exit calibration mode.
●
The external temperature sensor would now display 298 to 302.
Desoldering Gun Digital Temperature Calibration Example
●
The external temperature sensor displays 300 degrees.
●
The set temperature and displayed actual temperature of the
soldering iron is 350 degrees.
●
300 — 350 = -50. An additional adjustment of –50 degrees is
required. Upon entering calibration mode, the display shows “010”,
indicating a calibration number of 10 is already present.
●
Therefore 10-50= -50.
●
We adjust from “010” to “050-” by pressing the down button of the
Desoldering Gun Temperature Adjustment button (“5” from the
control panel)
●
Save and exit calibration mode.
●
The external temperature sensor would now display 298 to 302.
NOTES:
●
Calibration will only make the newly calibrated point the most
accurate. Other temperature points may be a little off.