DIGITAL CALIBRATION
J. Utilizing the Solder Iron Digital Temperature Calibration
By default, the system is properly calibrated but for some cases
when a little adjustment of the soldering iron temperature is required
the following procedure can be done.
1. Turn on the soldering iron function switch.
2. Set to appropriate temperature you want to calibrate. Place the tip
of the soldering iron on an external temperature meter.
3. The readings on the external temperature sensor should be more
or less equal to the displayed temperature.
4. If there are large discrepancy in the temperature reading we can
re-calibrate the temperature setting. First write down the set
temperature of the soldering iron and the actual temperature
reading from the external temperature meter. For example:
set temperature =350
external temperature = 300
5. Turn off the Soldering Iron Function switch (“14” from the control
panel) .Ensure that the Desoldering Gun Function switch is in the
off position(“12” from the control panel) .
6. Press and hold the UP button of the Desoldering Gun
Temperature Adjustment button (“5” from the control panel) .
7. The Soldering Iron Temperature Display (“11” from the control
panel) . Will switch to four zeros “000” indicating it is now in the
soldering iron calibration mode. Release the UP button of the
Desoldering Gun Temperature Adjustment button
8. Use the Soldering Iron Temperature Adjustment buttons (“4” from
the control panel) to increase or decrease the calibration values.
In our example the set temperature is 350 but the actual
temperature is 300, There is a difference of 50 degrees. Press the
up button until we reach “050 ”.
9. Confirm the change by pressing and holding the “UP” button of
the Desoldering Gun Temperature Adjustment button (“5” from
the control panel).