4-5 Calibration File Updates
Chapter 4 — Adjustment
4-4
PN: 10370-10372 Rev. B
MG3702xA MM
4-5
Calibration File Updates
The following table lists the adjustment procedures (detailed below), the corresponding calibration files that
get updated, and whether the adjustments are done manually or by the Levels Cal Station.
Step Attenuator, 20 GHz
Section 4-7 “Frequency Reference Adjustment”
Section 4-8 “Level Detector Tables and Level Flatness Adjustment”
Section 4-9 “Level SWF Band Adjustment”
Section 4-10 “Level ALC Adjustment”
Section 3-6 “Internal Time Base Aging Rate Test (Optional)”
Section 3-7 “Spurious Signals Tests”
Section 3-8 “SSB Phase Noise Test”
Section 3-9 “Power Level Accuracy and Flatness Tests”
Section 3-10 “Pulse Modulation Tests”
All Other RF Components
Section 4-7 “Frequency Reference Adjustment”
Section 4-8 “Level Detector Tables and Level Flatness Adjustment”
Section 4-9 “Level SWF Band Adjustment”
Section 4-10 “Level ALC Adjustment”
Section 3-6 “Internal Time Base Aging Rate Test (Optional)”
Section 3-7 “Spurious Signals Tests”
Section 3-8 “SSB Phase Noise Test”
Section 3-9 “Power Level Accuracy and Flatness Tests”
Section 3-10 “Pulse Modulation Tests”
Table 4-3.
Adjustment List (1 of 2)
Calibration Adjustments
Calibration Files
Updated by Manual Calibration or
Levels Calibration Station
Utility DVM Cal
Dvm.xml
Manual Calibration
Dvm.log
Manual Calibration
Frequency Reference Cal
FrequencyRefCal.log
Manual Calibration
Level Detector Tables Cal
DetectorTableLowGain.xml
Levels Calibration Station
DetectorTableMidGain.xml
Levels Calibration Station
DetectorTableHighGain.xml
Levels Calibration Station
Level Flatness Cal
FlatnessTable00.xml
Levels Calibration Station
FlatnessTable01.xml
Levels Calibration Station
FlatnessTable02.xml
Levels Calibration Station
FlatnessTable03.xml
Levels Calibration Station
FlatnessTable04.xml
Levels Calibration Station
FlatnessTable05.xml
Levels Calibration Station
FlatnessTable06.xml
Levels Calibration Station
FlatnessTable07.xml
Levels Calibration Station
FlatnessTable08.xml
Levels Calibration Station
FlatnessTable09.xml
Levels Calibration Station
FlatnessTable10.xml
Levels Calibration Station
FlatnessTable11.xml
Levels Calibration Station
Table 4-2.
Adjustment and Verifications Required after Subassembly Replacement (3 of 3)
Subassembly/RF Component Replaced Perform the Following Adjustments and Verifications
Содержание MG3702xA Series
Страница 4: ......
Страница 5: ......
Страница 8: ...TG 8 PN 10370 10372 Rev B MG3702xA MM ...
Страница 24: ...1 12 Test Equipment List Chapter 1 General Information 1 10 PN 10370 10372 Rev B MG3702xA MM ...
Страница 42: ...2 5 RF Deck Assemblies Chapter 2 Functional Description 2 18 PN 10370 10372 Rev B MG3702xA MM ...
Страница 120: ...5 5 Troubleshooting Tables Chapter 5 Troubleshooting 5 24 PN 10370 10372 Rev B MG3702xA MM ...
Страница 144: ...6 13 RF Components Chapter 6 Removal and Replacement Procedures 6 24 PN 10370 10372 Rev B MG3702xA MM ...
Страница 178: ...B 1 MG3702xA Technical Data Sheet Appendix B Performance Specifications B 2 PN 10370 10372 Rev B MG3702xA MM ...
Страница 182: ...Index 4 PN 10370 10372 Rev B MG3702xA MM ...
Страница 183: ......