3-9 Power Level Accuracy and Flatness Tests
Chapter 3 — Performance Verification
3-24
PN: 10370-10372 Rev. B
MG3702xA MM
Power Level Accuracy (Models without Option 002)
For the following test, refer to the test record table in
“Section 3-9 Power Level Accuracy and
,
“Power Level Accuracy (Models without Option 002)”
Power level accuracy for power levels of –50 dBm and above are tested by stepping the output power level down
in 5 dB increments from its maximum rated power level and measuring the output power level using a power
meter at each step.
Test Setup
For all power level flatness measurements connect the equipment, shown in
follows:
1.
Calibrate the power meter with the appropriate power sensor.
2.
Connect the power sensor to the RF Output of the MG3702xA (use a fixed attenuator when
measuring power levels above +19 dBm).
3.
Connect the special AUX I/O interface cable (Anritsu Part No. 806-97) to the MG3702xA rear panel
AUX I/O connector. Connect the cable BNC connectors as follows:
a.
Connect the cable labeled
SEQ SYNC
to the power meter rear panel
INPUT 1 DIGITAL
connector.
b.
Connect the cable labeled
HORIZ OUT
to the power meter rear panel
INPUT 2 ANALOG
connector.
4.
Set up the power meter as follows:
a.
Reset the power meter by pressing:
System
|
Setup
| -
more
- |
PRESET
|
RESET
b.
Configure the power meter to perform power measurements by pressing:
Sensor
|
Setup
|
MODE
Press
MODE
until
Default
is displayed in the
Measurement MODE
field.
c.
Configure the power sensor calibration factor source by pressing:
Sensor
|
CalFactor
|
SOURCE
Press
SOURCE
until
V/GHz
is displayed in the
Cal Factor SOURCE
field.
d.
Setup the minimum
V/GHz
range by pressing:
Setup
|
Start F
e.
Enter the minimum frequency of the MG3702xA under test or minimum of the power meter.
f.
Setup the maximum
V/GHz
range by pressing:
Setup
|
Stop F
g.
Enter the maximum frequency of the MG3702xA under test or maximum of the power
meter.
h.
Press any hard key to begin the measurement.
5.
Set up the MG3702xA as follows:
Caution
To prevent damage to the power sensor, use a fixed attenuator when measuring power levels above
+19 dBm.
Caution
Caution To prevent damage to the power sensor, use a fixed attenuator when measuring power
levels above +19 dBm.
Содержание MG3702xA Series
Страница 4: ......
Страница 5: ......
Страница 8: ...TG 8 PN 10370 10372 Rev B MG3702xA MM ...
Страница 24: ...1 12 Test Equipment List Chapter 1 General Information 1 10 PN 10370 10372 Rev B MG3702xA MM ...
Страница 42: ...2 5 RF Deck Assemblies Chapter 2 Functional Description 2 18 PN 10370 10372 Rev B MG3702xA MM ...
Страница 120: ...5 5 Troubleshooting Tables Chapter 5 Troubleshooting 5 24 PN 10370 10372 Rev B MG3702xA MM ...
Страница 144: ...6 13 RF Components Chapter 6 Removal and Replacement Procedures 6 24 PN 10370 10372 Rev B MG3702xA MM ...
Страница 178: ...B 1 MG3702xA Technical Data Sheet Appendix B Performance Specifications B 2 PN 10370 10372 Rev B MG3702xA MM ...
Страница 182: ...Index 4 PN 10370 10372 Rev B MG3702xA MM ...
Страница 183: ......