![Anritsu MG3702xA Series Скачать руководство пользователя страница 53](http://html.mh-extra.com/html/anritsu/mg3702xa-series/mg3702xa-series_maintenance-manual_2945427053.webp)
Chapter 3 — Performance Verification
3-8 SSB Phase Noise Test
MG3702xA MM
PN: 10370-10372 Rev. B
3-11
3-8
SSB Phase Noise Test
This section contains two different procedures depending on what test equipment is available:
•
“Procedure Using Agilent Equipment”
•
“Procedure Using AeroFlex Equipment”
Procedure Using Agilent Equipment
This section contains instructions to perform a SSB phase noise test using the Agilent E5052B Signal Source
Analyzer and E5053A Down Converter. For the following test, refer to the test record table in
,
“Section 3-8 SSB Phase Noise Test: Agilent Test Equipment”
Required Equipment
MG3702xA Test Setup
The MG3702xA is setup as follows:
1.
Preset the instrument by pressing
Preset
. The CW menu is displayed.
2.
Press
Edit Frequency
to open the current frequency parameter for editing.
3.
Set F0 to the CW frequency indicated in
“Section 3-8 SSB Phase Noise Test: Agilent
4.
Press
Edit Level
to open the current power level parameter for editing. Set
L0
to +10 dBm or the
maximum leveled power of the instrument, whichever is lower.
5.
Connect the MG3702xA as shown in
.
a.
Connect the output of the MG3702xA to the input of the splitter.
b.
Connect one of the outputs of the splitter to the RF input of the E5052B
c.
Connect one of the outputs of the splitter to the RF input of the down converter.
Table 3-2.
Recommended Equipment List
Part Number
Quantity
Description
Vendor
K222B
1
Insertable F-F
Anritsu
34NKF50
1
N Male to K Female Adapter
Anritsu
K241C
1
Power Splitter
Anritsu
SC3855
1
3670K50-2 KM-KM flex cable
Anritsu
E5052B
1
Signal Source Analyzer
Agilent
E5053A
1
Down Converter
Agilent
K120MM-20CM
2
Semi-rigid cables K120MM-20CM (this will need to be bent
on site to fit)
Anritsu
1
Mouse, Optical, PS2/USB for E5052B
1
Keyboard, USB for E5052B
Note
The MG3702xA, E5052B and the E5053A Down converter must be powered on for a minimum of
30 minutes before performing these measurements. The E5052B and E5053A must be installed and
set up according to the instructions supplied with the instruments before continuing with this
procedure.
Note
If you are not using the splitter, reduce L0 power level to +8 dBm.
Содержание MG3702xA Series
Страница 4: ......
Страница 5: ......
Страница 8: ...TG 8 PN 10370 10372 Rev B MG3702xA MM ...
Страница 24: ...1 12 Test Equipment List Chapter 1 General Information 1 10 PN 10370 10372 Rev B MG3702xA MM ...
Страница 42: ...2 5 RF Deck Assemblies Chapter 2 Functional Description 2 18 PN 10370 10372 Rev B MG3702xA MM ...
Страница 120: ...5 5 Troubleshooting Tables Chapter 5 Troubleshooting 5 24 PN 10370 10372 Rev B MG3702xA MM ...
Страница 144: ...6 13 RF Components Chapter 6 Removal and Replacement Procedures 6 24 PN 10370 10372 Rev B MG3702xA MM ...
Страница 178: ...B 1 MG3702xA Technical Data Sheet Appendix B Performance Specifications B 2 PN 10370 10372 Rev B MG3702xA MM ...
Страница 182: ...Index 4 PN 10370 10372 Rev B MG3702xA MM ...
Страница 183: ......