![Anritsu MG3702xA Series Скачать руководство пользователя страница 58](http://html.mh-extra.com/html/anritsu/mg3702xa-series/mg3702xa-series_maintenance-manual_2945427058.webp)
3-8 SSB Phase Noise Test
Chapter 3 — Performance Verification
3-16
PN: 10370-10372 Rev. B
MG3702xA MM
Procedure Using AeroFlex Equipment
This section contains instructions to perform a SSB phase noise test using the AeroFlex PN9000 phase noise
system. For the following test, refer to the test record table in
,
“Section 3-8 SSB Phase Noise Test:
.
The following test can be used to verify that the MG3702xA meets its single sideband phase noise
specifications. For this test, an Anritsu MG3694A or B (with Options 3 and 4) signal generator is required to
act as a local oscillator (LO).
lists the single side band requirements for the local oscillator.
The CW RF output of the MG3702xA is mixed with the CW RF output from the MG3694B LO, which is offset
by 100 MHz. Single sideband phase noise is measured at offsets of 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz, and
1 MHz away from the resultant 100 MHz IF.
Test Setup
The PN9000 software must be installed and set up according to the instructions supplied with the phase
noise measurement system before continuing with this procedure.
Connect the equipment, shown in
, as follows:
1.
Connect a GPIB interface cable from the PN9000 system to the MG3694B (LO) rear panel
IEEE-488 GPIB connector.
2.
Connect a GPIB interface cable from the PN9000 system to the MG3702xA (DUT) rear panel
IEEE-488 GPIB connector.
3.
Connect the MG3694B (LO) RF Output to the LO INPUT of the PN9000 phase detector module.
4.
Connect the MG3702xA (DUT) RF Output to the RF INPUT of the PN9000 phase detector module.
Table 3-3.
Single Side Band Requirements for Local Oscillator
Frequency Range
Offset from Carrier
10 Hz
100 Hz
1 kHz
10 kHz
100 kHz
1 MHz
10 MHz to
15.625 MHz
-105 -126 -139 -142 -141 -145
> 15.625 MHz to
31.25 MHz
-99 -120 -134 -137 -137 -145
> 31.25 MHz to
62.5 MHz
-90 -114 -129 -136 -136 -144
> 62.5 MHz to
125 MHz
-88 -108 -127 -135 -133 -144
> 125 MHz to
250 MHz
-84 -102 -125 -132 -130 -143
> 250 MHz to
500 MHz
-77 -99 -123 -125 -124 -142
> 500 MHz to
1050 MHz
-71 -93 -118 -121 -119 -138
> 1050 MHz to
2200 MHz
-66 -86 -112 -115 -113 -135
> 2.2 GHz to
6 GHz
-54 -77 -104 -108 -107 -130
> 6 GHz to
10 GHz
-52 -73 -100 -107 -107 -128
> 10 GHz to
20 GHz
-45 -68 -94 -102 -102 -125
> 20 GHz to
40 GHz
-45 -63 -92 -98 -98 -119
Note
Before performing this procedure ensure that all test equipment is calibrated. Refer to the
manufacturer's test equipment manual.
Содержание MG3702xA Series
Страница 4: ......
Страница 5: ......
Страница 8: ...TG 8 PN 10370 10372 Rev B MG3702xA MM ...
Страница 24: ...1 12 Test Equipment List Chapter 1 General Information 1 10 PN 10370 10372 Rev B MG3702xA MM ...
Страница 42: ...2 5 RF Deck Assemblies Chapter 2 Functional Description 2 18 PN 10370 10372 Rev B MG3702xA MM ...
Страница 120: ...5 5 Troubleshooting Tables Chapter 5 Troubleshooting 5 24 PN 10370 10372 Rev B MG3702xA MM ...
Страница 144: ...6 13 RF Components Chapter 6 Removal and Replacement Procedures 6 24 PN 10370 10372 Rev B MG3702xA MM ...
Страница 178: ...B 1 MG3702xA Technical Data Sheet Appendix B Performance Specifications B 2 PN 10370 10372 Rev B MG3702xA MM ...
Страница 182: ...Index 4 PN 10370 10372 Rev B MG3702xA MM ...