Operator’s Manual
pg. 34
Rev 05/12/21
Select Crude Fib
˅
<Enter>
Insert Samples
Close Lid<Enter>
Hot Water On
<Start>
<Start>
Check the Temperature Control using the Boil Test
To check the ANKOM
2000
Fiber Analyzer temperature control, do a Boil Test
once per month
as specified in the
procedure below.
1.
Disconnect cubitainers if connected.
2.
Fill a pitcher with approximately 2L of hot water (50 ˚C or above) and set aside for later.
3.
Turn on the instrument. Remove the bag suspender from the vessel.
on the display, press the
ENTER
button.
4.
When you see
on the display, press the
ENTER
button.
5.
When you see
(
Do not
insert samples.
Do not
close lid.)
on the display, press the
START
button.
6.
When you see
7.
Record Voltage:
a.
Record the voltage shown on the display before pouring in the water. _______V
(Should be between 5-10 VDC)
b.
Pour the 2L of hot water into the digestion vessel so that the
glass level sensor tip is under water.
c.
Record the voltage after pouring in the water. _______V
(Should be between 0-3 VDC)
Agitation and heat begin automatically.
8.
Do not close the lid. Occasionally stir the water to ensure uniform heating.
The water should begin to boil in 5-15 minutes.
9.
When a strong boil is reached, stir the water in the vessel and record:
a.
The temperature on the controller. _______˚C
b.
The temperature on a calibrated thermometer. _______˚C
(Place the thermometer near the probe at the bottom of the vessel.)
c.
The temperature shown on the display. _______˚C
If these three temperatures are not within ½ degree of each other, contact
ANKOM Technology for assistance. Send results to [email protected].
Level Sensor Tip
Controller
Calibrated
Thermometer
Temperature Probe
Содержание 2000
Страница 2: ...This page intentionally left blank ...
Страница 4: ...This page intentionally left blank ...
Страница 18: ...Operator s Manual pg 18 Rev 05 12 21 This page intentionally left blank ...
Страница 24: ...Operator s Manual pg 24 Rev 05 12 21 This page intentionally left blank ...
Страница 30: ...Operator s Manual pg 30 Rev 05 12 21 This page intentionally left blank ...
Страница 40: ...Operator s Manual pg 40 Rev 05 12 21 Appendix B Wiring Diagrams 1 of 4 ...
Страница 41: ...Operator s Manual Rev 05 12 21 pg 41 Appendix A Wiring Diagrams 2 of 4 ...
Страница 42: ...Operator s Manual pg 42 Rev 05 12 21 Appendix A Wiring Diagrams 3 of 4 ...
Страница 43: ...Operator s Manual Rev 05 12 21 pg 43 Appendix A Wiring Diagrams 4 of 4 ...