Operator’s Manual
pg. 22
Rev 05/12/21
11.
Press the arrow keys on the Keypad until you see “Select NDF” on
the Display.
The NDF analysis will use the following default settings:
•
75 minute digestion
•
four 5 minute rinses
If you want to run a custom NDF analysis that allows you to set the digestion
time and the number of rinses, press the arrow keys until you see “Select
Custom” on the Display.
12.
Press ENTER on the Keypad and follow the prompts on the Display to set up the instrument for NDF
analysis.
If you do not want to use the Cubitainers to automatically add your solutions,
you can manually fill the vessel for each procedure. First make sure that the
hoses to the Cubitainers are disconnected. Immediately after executing step 13
below, pour 2L of solution directly into the vessel. Please note that the
solution must cover the level sensor in order for the instrument to start its
operation. Continue the procedure with step 14.
13.
Press START on the Keypad. Solution from the Cubitainer will flow into the vessel through Port A.
Once the analysis begins, digestions and rinses occur automatically, with the
analyzer Display providing information about the process time remaining, the
temperature, and the pressure. Pressing STOP on the Keypad at any time
during the analysis ends the operation and opens the drain to exhaust the
solution.
14.
After the ND solution has been added and agitation begins, manually add 20g of Na
2
SO
3
and 4.0ml of
alpha-amylase directly into the Vessel.
15.
Close the Vessel Lid and tighten it by turning the Vessel Clamp Handle.
16.
When the “Extraction Done” message appears on the Display, the analyzer operation is complete. Open
the Vessel Lid, remove the bags and place them in an appropriately sized beaker.
17.
With your hands, gently press out excess water from the bags into the beaker
and pour off the water from the beaker.
18.
With the bags in the beaker, add enough acetone to cover them. Let the bags
soak in acetone for 3 – 5 minutes. Then pour off the acetone.
19.
With your hands, gently press out excess acetone from the bags into the
beaker and pour off the acetone from the beaker.
Select NDF
v
<Enter>
Содержание 2000
Страница 2: ...This page intentionally left blank ...
Страница 4: ...This page intentionally left blank ...
Страница 18: ...Operator s Manual pg 18 Rev 05 12 21 This page intentionally left blank ...
Страница 24: ...Operator s Manual pg 24 Rev 05 12 21 This page intentionally left blank ...
Страница 30: ...Operator s Manual pg 30 Rev 05 12 21 This page intentionally left blank ...
Страница 40: ...Operator s Manual pg 40 Rev 05 12 21 Appendix B Wiring Diagrams 1 of 4 ...
Страница 41: ...Operator s Manual Rev 05 12 21 pg 41 Appendix A Wiring Diagrams 2 of 4 ...
Страница 42: ...Operator s Manual pg 42 Rev 05 12 21 Appendix A Wiring Diagrams 3 of 4 ...
Страница 43: ...Operator s Manual Rev 05 12 21 pg 43 Appendix A Wiring Diagrams 4 of 4 ...