Operator’s Manual
pg. 20
Rev 05/12/21
9.
Seal the filter bag within 4mm of its open end. Keep the sealer arm down for 2 – 3 seconds after the red
sealer light turns off (to cool the seal). The seal can be seen as a solid melted stripe along the top edge of
the filter bag. If the seal is not strong, re-seal the bag.
10.
Spread the sample out uniformly within the filter bag by shaking and flicking the bag to eliminate
clumping.
11.
Repeat steps 5 – 10 for all filter bags that will be used in the Analyzer. (Up to 24 bags can be processed
during one procedure.)
If your samples contain soybean products or >5% fat
:
Before doing the NDF analysis in the ANKOM
2000
, you will need to do a pre-
extraction. For samples containing non-roasted soybean or >5% fat, follow
the pre-extraction steps below:
1.
Place the filter bags with sample (up to 23) into a container with a
top.
2.
Pour enough fresh acetone into the container to cover the bags.
3.
Put the top on the container.
4.
Shake the container 10 times and allow bags to soak for 10 minutes.
5.
Pour out and dispose of the acetone.
6.
Execute steps 1 through 5 a total of two times.
7.
Place the bags on a wire screen to air-dry.
If your samples contain roasted soybean:
Before doing the NDF analysis in the ANKOM
2000
, you will need to do a pre-
extraction. For samples containing roasted soybean, follow the pre-extraction
steps below:
1.
Place the filter bags with sample (up to 23) into a container with a
top.
2.
Pour enough fresh acetone into the container to cover the bags.
3.
Put the top on the container.
4.
Shake the container 10 times.
5.
Pour out and dispose of the acetone.
6.
Pour fresh acetone into the container and allow the samples to soak
for twelve hours.
7.
Pour out the acetone.
8.
Place the bags on a wire screen to air-dry.
12.
Place the filter bags with sample and at least one empty bag (used as a Blank) into
the Bag Suspender trays as shown (maximum of three bags per tray).
13.
Stack each tray on the Bag Suspender rod (eight trays in total) with each tray
rotated 120 degrees from the tray below.
You must use all eight trays even if they are empty.
14.
Add the 9
th
tray to the top of the Bag Suspender rod. This tray contains no filter bags and acts as a cover.
The samples are now ready for the NDF analysis procedure.
Содержание 2000
Страница 2: ...This page intentionally left blank ...
Страница 4: ...This page intentionally left blank ...
Страница 18: ...Operator s Manual pg 18 Rev 05 12 21 This page intentionally left blank ...
Страница 24: ...Operator s Manual pg 24 Rev 05 12 21 This page intentionally left blank ...
Страница 30: ...Operator s Manual pg 30 Rev 05 12 21 This page intentionally left blank ...
Страница 40: ...Operator s Manual pg 40 Rev 05 12 21 Appendix B Wiring Diagrams 1 of 4 ...
Страница 41: ...Operator s Manual Rev 05 12 21 pg 41 Appendix A Wiring Diagrams 2 of 4 ...
Страница 42: ...Operator s Manual pg 42 Rev 05 12 21 Appendix A Wiring Diagrams 3 of 4 ...
Страница 43: ...Operator s Manual Rev 05 12 21 pg 43 Appendix A Wiring Diagrams 4 of 4 ...