![Andor Technology Zyla sCMOS 4.2 PLUS Скачать руководство пользователя страница 70](http://html1.mh-extra.com/html/andor-technology/zyla-scmos-4-2-plus/zyla-scmos-4-2-plus_hardware-manual_2942422070.webp)
version 2.0 rev 18 Feb 2021
70
6.6 Fire, Aux_Out and Arm outputs not functioning correctly
These are 5 V TTL outputs which should not be used to drive low impedance loads. Please refer to
Section
1.7.2
and
1.7.3
for suggested terminations.
6.7 External Trigger input not functioning correctly
This is a 5 V TTL input which should be driven from a 5V TTL compatible source. Please refer to
Section 1.7.2
and
1.7.3
for further details.
6.8 Condensation Appearing On Exterior Of Camera (Water Cooled
Model)
NEVER USE WATER THAT HAS BEEN CHILLED BELOW THE DEW POINT OF THE AMBIENT ENVIRONMENT TO
COOL THE CAMERA.
You may see condensation on the outside of the camera body if the cooling water is at too low a temperature
or if the water flow is too high. The first signs of condensation will usually be visible around the connectors
where the water tubes are attached. If this occurs carry out the following actions:
1.
Switch off the system
2. Wipe the camera with a soft, dry cloth.
NOTE: It is likely there will already be condensation on the cooling block and cooling fins inside the
camera.
3. Set the camera aside to dry for several hours before you attempt reuse.
4. Before reuse blow dry gas through the cooling slits on the side of the camera to remove any residual
moisture.
Use warmer water or reduce the flow of water when you start using the device again.
NOTE: This is not an issue when using a Re-circulator which eliminates the dew point problem.
Refer to
Appendix B
for a Dew Point Graph.
Содержание Zyla sCMOS 4.2 PLUS