![Andor Technology Zyla sCMOS 4.2 PLUS Скачать руководство пользователя страница 32](http://html1.mh-extra.com/html/andor-technology/zyla-scmos-4-2-plus/zyla-scmos-4-2-plus_hardware-manual_2942422032.webp)
version 2.0 rev 18 Feb 2021
32
2.6.2.3 Rolling Shutter Internal Triggering (Non-Overlap Mode)
Internal Trigger Mode allows the user to configure an exposure time and cycle time. For Internal Triggering
Non-overlap mode, the exact acquisition sequence depends on the exposure time and cycle time set as shown
in
Figure 16
. The following diagrams show the behaviour of TTL outputs
‘Fire
’,
Fire n
’, ‘
Fire ALL
’ and ‘
Fire ANY
’.
Fire ALL
indicates the time period within a frame during which all rows are exposing simultaneously.
Fire ANY
indicates the time period within a frame during which any row is exposing
Initially, the entire sensor is held in a global clear state to ensure that there is no charge build-up on the sensor.
Global Clear goes LOW and a frame read out is initiated. This frame is discarded as it does not contain the
correct exposure period. Reading out a frame effectively begins a new exposure. When the exposure period
has completed, a signal frame read out phase begins. When the frame has been read out completely, the
Global Clear is held HIGH until the user-defined cycle time is achieved.
Figure 16: Rolling Shutter Internal Triggering
Table 12:
Rolling Shutter Internal Triggering (Long Exposures) Timing Parameters
Parameter
Minimum
Maximum
Exposure
1 Row
30 s
Cycle Time (1/Frame Rate)
Ex 1 Frame + 1 Row
20,000 s
Acquisition Start Delay
-
1 Row
Содержание Zyla sCMOS 4.2 PLUS