50
2015
“What
to
do
if...“
8.3
Fault Removal Instructions
8.3.1
Replacement of Inlet Fuse(s)
Fig.
16: Fuses (two) in combi seal at the back of the pipettor
Replacement of system fusing
Fuse replacement is complete.
WARNING
There
is
danger
of
exposure
to
electric
voltage
!
Always
detach
the
power
supply
cable
from
the
line
power
socket
before
you
replace
a
fuse.
Check
for
absense
of
voltage
in
the
pipettor.
Description
Type/Rated specs
Line
power
fuse
T 6,3 A,
250V
AC,
5x20 mm
1.
Turn
power
supply
off.
2.
Detach
line
power
cord
from
system
inlet
port
and
mains
socket.
3.
Use
a
small
screwdriver
to
extract
the
fuse
holder
from
the
combi
seal
4.
Perform
check
to
identify
the
defective
fuse.
5.
Remove
defective
fuse
from
fuse
holder.
6.
Install
replacement
fuse
(T6.3A)
in
fuse
holder.
7.
Restore
the
fuse
holder
to
its
position.
8.
Turn
power
supply
on
again.
Содержание 845-00003-03
Страница 1: ...InnuPure C96 User Manual ...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 7: ...2015 VII ...
Страница 8: ...VIII 2015 ...
Страница 12: ...XII 2015 ...
Страница 20: ...8 2015 General Information ...
Страница 44: ...32 2015 Technical Description ...
Страница 54: ...42 2015 Routine Start Up Procedure ...
Страница 60: ...48 2015 Operation ...
Страница 64: ...52 2015 What to do if ...
Страница 72: ...60 2015 Shutting Down ...
Страница 76: ...64 2015 Waste Disposal ...