Conductivity Electrode Care and Maintenance
●
Ensure that the conductivity electrode is thoroughly washed with distilled water after each use.
●
If there is a build-up of solids inside the measurement area of the cell, these should be removed very carefully with a cotton bud soaked in solvent,
taking care not to touch the metal parts of the inner cell.
DO Probe Care and Maintenance
●
Please always keep probe’s membrane moist.
●
If you do not use the dissolved oxygen probe for long periods, please screw off membrane cap and rinse the probe's cathode, anode and the
membrane with deionized water, then soak up residual water on them with filter paper, and install the probe.
Troubleshooting
LCD DISPLAY
CAUSE
CORRECTIVE ACTION
Electrode dried out
Soak the pH electrode in 3M KCL solution for 15 minutes
Soak the Ion Selective Electrode in the diluted standard solution
Soak the conductivity electrode in tap water for a few minutes
Measured value is out of range
Check the electrode whether clogged, dirty or broken
Incorrect calibration solutions
Using the fresh calibration solutions for calibration
Setting value does not match calibration solution
Reset the calibration value
DO electrolyte solution is depleted
Refilling electrolyte solution
Zero oxygen solution is contaminated
Replace the calibration solution
pH electrode is out of service life
Replace the pH electrode
Keypad is not working properly
Replace the batteries
Cathode and Anode
Membrane cap
EC910 Portable Multiparameter Water Quality Meter