14
6.
Press
or
key to set the date and time.
7.
Press ENTER key to confirm, the meter returns to measurement mode. Setting is completed.
Temperature Compensation
In order to get accurate measuring results, you need to enable the manual or automatic temperature compensation before measurement or calibration.
AUTOMATIC TEMPERATURE COMPENSATION:
●
Insert the connector of temperature probe into the meter’s phone jack.
●
The “ATC” indicator will show on the display, the meter is now switched to automatic temperature compensation mode.
MANUAL TEMPERATURE COMPENSATION:
1.
DO NOT connect the temperature probe to meter.
2.
Press and hold the
o
C key for 3 seconds to enter temperature setting mode.
3.
Press
or
key to set the temperature value of sample.
4.
Press ENTER key to confirm, the meter returns to measurement mode. Setting is completed.
In the temperature setting mode, press
or
key once, the setting value will increase or decrease by 0.1. Press and hold the
or
key, the
setting value will increase or decrease by 1.
In the dissolved oxygen mode, DO NOT connect the temperature probe to meter. The dissolved oxygen probe has a built-in temperature sensor.
EC910 Portable Multiparameter Water Quality Meter