35
1.5 Dip the dissolved oxygen probe into the air-saturated water for 3 to 5 minutes.
1.6 Press ENTER key to confirm, “Calibration” indicator begins flashing.
1.7 Wait for the measured value to stabilize, the display will show “END”. The meter returns to measurement mode automatically. Single point calibration
is completed.
2 POINTS CALIBRATION:
2.1 Make sure that you have selected 2 points calibration in the setup menu.
2.2 Press CAL key, the meter shows “8.25mg/L/CAL1” (@25
o
C).
2.3 Press
or
key until the display shows “0.00mg/L/CAL1”.
Enter
MI
EC910 Portable Multiparameter Water Quality Meter