24
8.
Rinse the Ion Selective Electrode with distilled water. Dip the electrode into corresponding calibration solution. Stir the sensor gently.
9.
Press ENTER key, “Calibration” indicator begins flashing. Wait for the measured value to stabilize, the display shows “1/CAL3”. The meter prompts
you to continue with third point calibration.
10. Repeat steps 8 to 9 above until the display shows “END”, the meter returns to measurement mode automatically. Calibration is completed.
EXIT THE CALIBRATION:
During the setup mode, press MEAS key, the meter returns to measurement mode immediately.
Enter
EC910 Portable Multiparameter Water Quality Meter