Conductivity Calibration
EC910 portable multiparameter water quality meter is capable of using the three types of conductivity electrodes. Before calibration or measurement,
you need to select the different sensors as per your sample concentration. The table below shows available sensors and its effective measuring ranges.
CONDUCTIVITY ELECTRODE
MEASURING RANGES
CELL CONSTANT
CON-0.1
0.1~100µS/cm
K=0.1
CON-1
0.01~10mS/cm
K=1
CON-10
0.1~200mS/cm
K=10
Make sure that you selected the cell constant between 0.1, 1 or 10 in the setup menu. If the current option is “USER”, the automatic calibration function
will be disabled (Refer to Setup Menu section).
In the automatic calibration mode, the meter allows up to 5 points calibration. To ensure higher accuracy, we recommend that you perform 3 points
calibration or select a standard value close to the sample value you are measuring. The meter will automatically detect these conductivity standard solutions
and prompt the user to calibrate the meter. When the calibration is completed, all new calibration values will automatically override existing data.
The following table shows acceptable conductivity range of calibration solution for each measuring range.
MEASURING RANGE
CALIBRATION SOLUTION RANGE
DEFAULT
0~20µS/cm
7~17µS/cm
10µS/cm
20~200µS/cm
70~170µS/cm
84µS/cm
200~2000µS/cm
700~1700µS/cm
1413µS/cm
2~20mS/cm
7~17mS/cm
12.88mS/cm
20~200mS/cm
70~170mS/cm
111.8mS/cm
SINGLE POINT CALIBRATION:
1.1 Press MODE key until the meter shows
indicator.
1.2 Rinse the conductivity electrode with distilled water, then rinse with a small amount of calibration solution.
1.3 Press CAL key, the meter enters the calibration mode.